Description
Design
EN 837-1
Nominal size in mm
160
Accuracy class
0.6
Scale ranges
0……0.6 to 0…..1.600 bar
or all other equivalent vacuum or combined pressure and vacuum ranges
Adjusment medium
Steady : Full scale value
Fluctuating : 0.9 x full scale value
Shoty time : 1.3 x full scale value
Permissible temperature
Ambient : -40….+60°C witout liquid filling
-20….+60°C with glycerine filling 1)
Medium : +20°C maximum without liquid filling
+100 °C maximum with liquid filling 1)
Temperature effect
When the temperature of the measuring system deviates from the reference temperature (+20°C) : max. 0.4%/10 K of full scale value
Application
- When liquid-filled case for applications with dynamic pressure loads or vibration
- For gaseous and liquid aggresive media that are not highly viscous or crystallising, also in aggresive ambience
- Precision measurement in laboratories
- High accuracy pressure measurement
- Testing industrial type pressure gauges
Special Features
- All stainless steel construction
- Knife edge pointer for optimana accuracy of reading
- Wear resistant precious novement from stainless steel
- Scale ranges up to 0….1.600 mbar