Description
Design :
DIN 16002
Nominal size in mm
100, 160
Accuracy class
Model 532.52 : 1.0
Model 532.53 : 1.6
Model 532.54 : 2.5
The measurement accuracy is ensured for ambient pressure fluctuations between 955 and 1.065 mbar (min. and max of atmosperic pressure)
Scale ranges
0…25 mbar to 0…25 bar absolute pressure
Pressure limitation
Steady : full scale value
Fluctuating : 0.9 x ful scale value
Overload safety
Minimum 1 bar absolute pressure (atmospheric pressure) in addition 10 x full scale value, max 25 bar absolute pressure
Permissible temperature
Ambient : -20….+60°C
Medium : +100°C maximum
Temperature effect
When the temperature of the measuring system deviates from the reference temperature (+20°C) : max ±0.8%/10 K of full value
Application
- Pressure measurement independent of fluctations in the atmospheric pressure
- For gaseous, liquid and aggresive media, also in aggresive environments
- Monitoring of vacuum pumps
- Control of vacuum packing machines
- Monitoring of condensation pressures and determination of vapour pressure in liquids
Special Features :
- High overload safety
- Long service life due to metaic media chamber selaing
- Media chamber protect ed against unauthorised intervention, DT-GM 86 08 176